Infrared Absorption Spectroscopy
Purpose: The purpose of this experiment is to provide experience with the methods commonly used to obtain infrared spectra of solids and liquids. These techniques may be used for either qualitative or quantitative application. Quantitation in infrared spectroscopy is considerably more difficult than in UV-Visible absorption and emission. Therefore, while quantitation is attempted here to introduce and illustrate some of the difficulties, the application of IR for compound identification is emphasized.
For this purpose an infrared spectrum is used in two ways. It may be interpreted to provide clues to the structure of an unknown. This approach is rarely conclusive, and must be supplemented with other types of evidence. Infrared spectra may also be used to compare a known compound with an unknown. For a conclusive identification, all features of both spectra that are more intense than instrument noise must match.
References: Bauman, Absorption Spectroscopy, Silverstein and Bassler, Spectrometric Identification of Organic Compounds
Equipment: Fourier transform infrared spectrophotometer (FTIR); diffuse reflectance and attenuated total reflectance (ATR) sampling accessories; mortar and pestle; rubber or plastic gloves; micro spatula; forceps; desiccators for cells and small pieces of equipment; 2 or 5 mL volumetric flask.
Reagents: IR grade KBr; dry chloroform, n-butanol; solid & liquid samples.
Procedure
All auxiliary equipment such as mortar, pestle, and diffuse reflectance sample cups should be washed, dried and stored in the desiccator at the end of each lab period. The ATR crystal should be carefully wiped with acetone. KBr should be stored in a drying oven when not immediately needed and otherwise in a desiccator.
1. Liquid Sample: The spectrum of your liquid sample is to be obtained using the ATR sampling accessory: on the neat liquid with the dry cell as a blank, and on a CHCl3 solution with neat CHCl3 as the blank. Follow the ATR instructions for placing the accessory in the FT-IR sample compartment and for placing the samples in the cell.
2. Effect of Impurities: Record the spectrum of pure n-butanol and 95% n-butanol, both neat, in the ATR cell.
3. Solid Sample: The solid spectra are obtained using the Diffuse Reflectance sampling accessory. Follow the sampling accessory instructions for placing the accessory in the FT-IR sample compartment and for loading samples into the sample cups. Samples are ground and "diluted" with KBr powder, with pure KBr powder used as the blank. Spectral distortions are minimized by keeping the powder granule size of the KBr blank, KBr "diluent", and sample roughly the same. (i.e., don't grind the heck out of the sample plus diluent but not the blank; grind both so that the final powder grain size is roughly similar. Also, don't grind excessively; this promotes water absorption by the sample and the KBr.)
Report
Liquid Sample -
Effect of Impurities -
Compare your spectrum of pure n-butanol with that of the
reference and note any discrepancies. Then compare the spectrum
of pure n-butanol with that of the 9% sample. Mark the
differences. If properly prepared, they will be minor; however,
even these small deviations are sufficient to vitiate an
identification if based entirely upon this evidence. Further
purification would be necessary for a conclusive identification.
Solid Sample -
Compare the spectrum with the reference (Sadtler index). Identify
any peaks that do not match. Determine whether they can be accounted
for by the KBr blank, or by the use of diffuse reflectance as the
sampling method.