Atomic Absorption
Spectrometry
Background: Atomic absorption
spectrometry involves the absorption of radiant energy by neutral
atoms in the gaseous state. The element being determined must be
reduced to the elemental state, vaporized, and imposed in the
beam of radiation from the source. This process is most
frequently accomplished by drawing a solution of the sample, as a
fine mist, into a flame. The flame thus serves a function
analogous to that of the cell and solution in conventional
absorption spectroscopy. The light beam from the source is passed
directly through the flame.
Quantitative analysis is easily performed by
measuring the radiation absorbed by the sample in the flame.
Typically a calibration curve is first prepared by measuring the
absorbance of a series of solutions of known concentration. The
sample(s) is then measured under same conditions. Excellent
results are generally obtained. About 65 elements can be
determined by atomic absorption with sensitivities down to 0.1
µg/mL in same cases.
Reference: Skoog, Holler and Nieman Principles
of Instrumental Analysis
Equipment: Buck model 200A atomic
absorption spectrophotometer, Dynac centrifuge, centrifuge tubes.
Reagents: 1.4 N HCl, 100 ppm
Stock Solution of Copper (made by diluting the appropriate weight
of copper sulfate pentahydrate with 1.4 N HCl), 20% (v/v)
trichloroacetic acid.
Unknown: Student's will choose there
own unknowns. Obtain samples that will have relatively low, yet
measurable copper concentrations. The procedure given below will
work for almost any sample. Biological or environmental samples
are usually the most interesting. Consult with the laboratory
Teaching Assistant or Instructor for details and/or procedures.
Procedure
Begin the digestion of your real sample. There
are two 30 minute periods of incubation during which time you
will continue with the instrument startup procedure below.
Sample and Standards
- Make 10, 20, and 40 ppm Cu standards from
the stock solution by diluting with 1.4 N HCl.
- To five labeled centrifuge tubs,
add 2.0 mL of the following solutions: 0 (Blank), 20, 40,
50, ppm Cu Standards. To the fifth tube add 2.0 mL of
unknown sample. Add 2.0 mL of 1.4 N HCl to all
tubes and mix. Allow tubes to sit for 30 minutes at room
temperature.
- Add 2.0 mL of 20% trichloracetic Acid
(TCA) to ALL tubes and allow to sit for an additional 20
minutes.
- Centrifuge all tubes for 10 minutes at
2000 RPM. (Use setting of 80 on Dynac Centrifuge to
provide 2000 RPM)
- Pour off supernatent into labeled clean
centrifuge tubes.
Measurements
See your instructor, he or she will demonstrate
the operation of the Buck 200A Atomic Absorption
Spectrophotometer.
- Adjust the gas flow and light the flame
with the assistance of the instructor according to the
instructions listed under Flame Types and Lighting
Instructions. DO NOT LIGHT THE FLAME UNTIL YOU ARE READY
TO ASSAY YOUR TREATED SAMPLES AND STANDARDS.
- Be sure the instrument is set to the
primary emission line of Copper (~324 nm). Use only the
minimum volume of sample necessary to achieve a stable
absorbance value (usually only 10 seconds aspiration time
is required).
- Adjust the meter to read zero absorbance
while aspirating the Blank solution. Next, read each
solution while aspirating it into the flame. It may be
necessary to repeat the measurements several times to
obtain reproducible values. It is also advised that the
zero absorbance setting be rechecked every 3-4 samples
for drift.
- Take readings of absorbance of each
solution, beginning with the tube 'Blank'.
- Return the instrument to the secondary
emission line as indicated by the COPPER INFORMATION
SHEET. Once again, set the zero absorbance level and read
the absorbance of each of the copper samples.
- Adjust the meter to read zero absorbance
while aspirating the Blank solution. Next, read each
solution while aspirating it into the flame. It may be
necessary to repeat the measurements several times to
obtain reproducible values. It is also advised that the
zero absorbance setting be rechecked every 3-4 samples
for drift.
- After your samples have been measured to
give sufficient reliability, allow the flame to aspirate
distilled water for 5-10 minutes to wash out the burner
head of any salts or particulate material.
- Finally, shut the burner off with
assistance from the instructor. Then, turn off the
electronics with the main power switch.
Questions
- What was the Cu concentration in the
unknown sample?
- Was the standard curve you used to
calculate the Cu concentration in the unknown sample
adequate? If not, how would you have changed it?
- Why did you add trichloroacetic acid to
the sample?
- Were there any sources of interference? If
so, what were they?